WWW.MEASURETEK.NET PS-103 User/Technical Manual Contents subject to change without notice Version 1.0 08/2013
6. Configuration Parameters Setup 1. When the scale is off, press and hold ZERO/ON/OFF and PRINT/HOLD keys until ‘C0NF’ is shown, which indicates t
Para- meter Default Setting 12R975setting12R976 setting 12R977 setting 12R978setting12R979settingC8 2 Zero range for ZERO/ON/OFF button: 0 –power-on
Table1: use Kg as calibration unit: Display division value in different weight unit that can be used Calibration division value kg lb lb:oz 0.0001k
4. More Information for Configuration Parameters: 4.1 The division (d) of scale is determined by C3,C4 and C5 : If C3=3 (interval is 5), C4=2(deci
7. User Parameters Setup 1. In normal weighing mode, press UNIT and ZERO/ON/OFF until “USer” is shown to enter in the user parameters setup mode. 2
4. More Information for User Parameters Setting: U5 to set serial communication output format: (1). U5=0: No serial communication function. It will
weight data is filled by ‘_’). (3). If the zero point has an error, it will return ten ‘_’ characters. (4). The character will be ‘-’ for negative
(6). Command: L<CR> (4ch 0dh) Response: <LF> H1H2H3<CR><ETX> If HOLD function is enabled, go to or exit from HOLD mode, simi
8. Symbol Definitions: 0﹉﹉﹉ -Zero point is over the setting range 0﹍﹍﹍ -Zero is below the setting range Ad﹉﹉﹉ -Analog digital converter chip over ma
SYMPTOM PROBABLE CAUSE REMEDY EEP.E1 CONFIG parameters are not correctly set Re-set CONFIG parameters as technical manual instructed. EEP.E2 USER
TABLE OF CONTENTS 1. ... 1 INTROD
PS‐103 www.measuretek.net .1810.Replacement Parts Part Number Description MH12R97602G AC120V/DC6V 500mA UL adaptor 11.One
1. INTRODUCTION and INSTALLATION General and Safety Information Risk of Electrical Shock: Disconnect all power sources before making cable connect
Contents Indicator Scale Platform AC120V/DC6V 500mA UL Adapter 6’ USB Cable Owner’s Manual Outer Hexagonal Wrench Load Cell Wiring 1
2. OVERVIEW OF CONTROLS AND FUNCTIONS Indicator Display Character Definitions PS‐103 www.measuretek.net .3
Indicator Display >0<- Scale is zeroed and gross weight is 0, tare is 0. Net - Display reading is net weight; tare is not 0. lb, kg, o
3. OPERATIONS Normal Weighing Mode 1. Power on the scale by pressing the ZERO/ON/OFF key. 2. When the display stabilizes, but it doesn’t show zero
4. Calibration Note: (1) Before calibrating the scale, you should prepare standard weights (more than 10% of FS weight) for calibration. (2) In t
5. View ADC Code or Power Voltage 1. In normal weighing mode, press and hold HOLD/PRINT and ZERO/ON/OFF keys until “code” is shown, which means the